(Multi-User Model)Medi+Sure Canada Inc. 31A-328 Speedvale Ave. East, Suite 160, Guelph , Ontario, N1E 0J4, Canada.www.medisure.caToll Free: 1-855-MDI
8 Medi+Sure® Blood Glucose Test Strip How to determine whether blood sample is sufficient NOTE: If the sample is sufficient, the sampling window
9 HOW YOUR SYSTEM WORKS Installing or Replacing the Battery Your Medi+Sure® Blood Glucose Monitoring System is packed with one 3-volt lithium coin ba
10 10. Press the button. The hour flashes. 11. Press the button or button until the correct hour is displayed. 12. Press the button. The minute
11 Reviewing Results in Memory The Medi+Sure® Blood Glucose Meter automatically stores up to 960 test results with date and time. You can review test
12 14. To delete all the results in the memory, press the button for at least 3 seconds. “----” message will be shown on the screen. This indicates
13 3. Press the button and the temperature, and thermometer symbol will be shown on the screen for 3 seconds and the meter will turn off automat
14 BLOOD SAMPLING AND TESTING Finger Tip Blood Sampling 1. Wash hands in warm, soapy water. Rinse well and dry thoroughly. 2. Press a single use lanci
15 automatically. You will hear a beep letting you know the countdown has begun. 7. Look at the sampling window on the test strip. If you have enou
16 Testing Control Samples Prepare the test strip and Glucose Control Solutions. 1. Press the button. The meter will power on. 2. Insert Medi+Sure®
17 HOW TO CLEAN AND DISINFECT YOUR DEVICE STEP 1: Clean Before Disinfecting 1. Dispense CaviWipes towelette. 2. Clean surface of your device with
Medi+Sure® Multi-Patient User Guide - Table of Contents INTRODUCTION ...
18 4. If the meter is being operated by a second person who is providing testing assistance to the user, the meter should be disinfected prior to use
19 Information of CaviWipe disinfectant CaviWipes Disinfecting Towelettes are purchased from METREX RESEARCH CORP, and the EPA Reg. is 46781-8. CaviWi
20 IMPORTANT NOTES Caring For Your Meter Taking care of your meter is important. Please follow the guidelines shown below: 1. Do not disassemble your
21 No Memory in the Meter Message: If there is no memory in the meter, you press button two times and the meter shows “----”. Sample Application Mes
22 HI Message: If the test result is higher than 41.7 mmol /L the “HI” message is displayed. Lo Message: If the test result is lower than 1.1 mm
23 ERROR MESSAGES AND TROUBLESHOOTING When you operate the meter incorrectly, error messages are displayed. Each message is defined below. Cause: Te
24 Cause: The system is operating outside the appropriate temperature range 10°~40°C. Action: Move your meter and test strips to a place where th
25 EXPECTED VALUES1 Expected blood glucose levels for people without diabetes: Time Range (mmol/L) Fasting < 7.0 Two Hours After Meal < 11.1 1
26 METER WARRANTY Medi+Sure Canada Inc. warrants, to the original purchasers of the Medi+Sure® Blood Glucose Meter, that your meter shall be free of d
27 Medi+Sure® Log Book DATE TIME TEST RESULT AC / PC NOTES/Patient Name AC PC AC PC AC PC AC PC AC PC AC PC A
1 INTRODUCTION We would like to thank you for making a smart decision. The Medi+Sure® Blood Glucose Monitoring System is designed for Canadians. The
28 NOTES:
29 Personal Information Name Address Home Phone Mobile Phone Hospital Physician Email Diagnosis Date Additional Information:
2 TESTING PRINCIPLE The Medi+Sure® Blood Glucose Monitoring System is designed to provide an easy, accurate method for determining capillary blood glu
3 INDICATIONS FOR USE The Medi+Sure® Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary
4 IMPORTANT SAFETY INSTRUCTIONS Users must adhere to Standard Precautions when handling or using this device. All parts of the glucose monitoring syst
5 CONTENTS AND SPECIFICATIONS Kit Contents Medi+Sure® Multi-Patient Blood Glucose Meter 3-Volt Lithium Coin Battery (CR2032) User's Gui
6 System Operating Conditions Test Strips Use only Medi+Sure Blood Glucose Test Strips Operating Conditions Temperature: 10°~40° C Humidity: RH ≦
7 ABOUT THE Medi+Sure® SYSTEM Medi+Sure® LCD Display Show Blood Glucose Test Results and Messages Medi+Sure® Meter Back
Komentáře k této Příručce